How to Read a Certificate of Analysis (COA)
By Vanta Labs Research teamPublished 4 min read
What each part of a COA tells you, using the reports we publish as the example, what they do not cover, and how to check one with the laboratory.
A Certificate of Analysis (COA) is a laboratory's report on one specific batch of material — a report number, an access code, a measured purity figure, and what was and wasn't tested for. Here's what each part of ours actually means, using the Kovera Labs layout we publish.
Report number and access code
Each report carries a report number, for example KVR-2026-6B1171, and an access code printed beside it. Enter both at koveralabs.com/verify to check the report with the laboratory itself, rather than relying on our copy.
Sample information
The top of the report identifies what was tested: the product name, its form (lyophilized powder or liquid), the batch number and the labeled quantity of the vial, such as 5 mg or 10 mg. Each vial size we have had tested was tested as its own sample and has its own report.
Purity
Purity is measured by reversed-phase high-performance liquid chromatography (RP-HPLC). Our reports name the column (C18) and the detector: a diode-array detector (DAD) reading at 214 nm, or at 260 nm for NAD+.
Specification and measured figure
Each line of the test table sets a specification beside a measured figure. For purity, the specification printed on our reports is greater than 98%, and the figure beside it is what the laboratory measured in that sample, with a mark showing whether it met the specification. The specification is the pass mark; the measured figure is the finding. A product page on this site quotes the measured figure, and only for the vial size the report covers.
The chromatogram
The chromatogram is the trace the detector recorded as the sample passed through the column: time runs along the bottom, detector response up the side. The compound appears as the main peak; anything else the detector saw appears as smaller peaks. The purity figure comes from the areas under those peaks, which is why the method printed beside it matters (see Purity & Third-Party Testing, Explained).
Net content
Beside the labeled quantity, the report states the amount measured in the vial, so the two can be compared.
Identity
Identity is confirmed by liquid chromatography–mass spectrometry (LC-MS), which compares the mass measured in the sample with the compound's expected mass. Identity and purity answer different questions. A sample can be the right compound and still carry impurities, or be free of impurities and be the wrong compound, which is why a useful COA reports both.
Blends
For a blend such as GLOW or KLOW, the report gives an average purity and an LC-MS identification for the blend as a whole, and lists each component with its labeled and measured amount.
Liquid identity reports
Some liquids, such as B12, are covered by a different kind of report: an identity screen. It lists the compounds that LC-MS detected in the vial. As the report itself states, it establishes presence, not content or proportion, and it gives no purity figure. A compound missing from the list is not thereby shown to be absent: the screen only sees what its panel and method can detect.
Checking a report, step by step
- Match the product name and the labeled quantity to the vial you are looking at. A report for a different vial size does not describe yours.
- Match the batch number on the report to the one on your vial.
- Note the certified date. A report describes the sample as of that date.
- Enter the report number and access code at koveralabs.com/verify.
- Read the method beside each figure, so you know how it was obtained.
What our reports do not cover
None of the reports we have published tests for sterility, endotoxins, residual solvents, heavy metals or water content. A COA that does not report a test says nothing either way about it, and no page on this site should be read as if it did.
Where to find ours
Every published report is in our COA Library, and each product page links the report for the vial size selected, under Batch documentation, with the report number and access code. Products without a published report say so on the product page. How our reports are produced and shown is set out on Testing & quality.